Previous Page  4 / 8 Next Page
Information
Show Menu
Previous Page 4 / 8 Next Page
Page Background

Acura®

manual

835 Macropipettes

Instruments combining high-tech materials with user-friendly

ergonomics. Optimal for indoor and outdoor environmental analysis,

clinical chemistry and cell culture applications, they offer superior

performance and make pipetting safer than ever.

Product Advantages

Smart and reliable volume adjustment

Single-handed volume setting

Interchangeable nozzle protection filter

Justip™ adjustable tip ejector* fitting most tips

Swift-set user calibration system*

Shock, UV-light and autoclaving resistance

Pasteur pipette fitting 2 and 5mL models

* Socorex patented

Display Always Visible

Located on the front of the instrument, the display window allows

set volume to be visible at all times during pipetting. A comforting

reassurance for the user.

Smart and Reliable Volume Adjustment

State-of-the-art micrometric volume setting is performed smoothly and

precisely by turning push button. Finely machined click-stops and free

rotating cap prevent any unwanted volume alteration.

Patented Tip Ejection System

Ergonomically positioned, the large surface, soft padded ejector provides

for low-pressure activation. In addition, the shaft height adjustment

system called Justip™ allows optimal distance setting between ejector

and tip.

Pasteur Pipette Adapter

Adapter nozzles allow fitting of Pasteur pipette on Acura®

manual

2 and

5mL models in addition to polypropylene tips. The adapters have two

O-rings guaranteeing proper holding and tightness of standard 2mL

Pasteur pipettes (Ø 7mm).

Nozzle Protection Filter

Protection against liquid overflow and contamination.

Inaccuracy (E%)

Imprecision (CV%)

Cat. No.

Volume

Division

Min. Vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Tip Style

W810315

0.2 - 2mL

0.002mL

< ± 1.5 % < ± 1.0 % < ± 0.5 %

< 0.5 % < 0.3 % < 0.2 %

2mL

W810316

0.5 - 5mL

0.01mL

< ± 1.5 % < ± 1.1 % < ± 0.6 %

< 0.6 % < 0.5 % < 0.3 %

5mL

W810318

1 - 10mL

0.01mL

< ± 1.5 % < ± 0.7 % < ± 0.5 %

< 0.5 % < 0.3 % < 0.2 %

10mL

Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20°C and 25°C in accordance with ISO 8655

.

Performance and Ordering Information - Acura

®

manual

835 Macropipettes

Fixed Volume Pipettes available at

www.wheaton.com

Acura®

manual

Pipettes

835

0.2 - 2mL

0.5 - 5mL

1 - 10mL