Acura

®

manual

835 Macropipettes

Instruments combining high-tech materials with user-friendly ergonomics. Optimal

for indoor and outdoor environmental analysis, clinical chemistry and cell culture

applications, they offer superior performance and make pipetting safer than ever.

Product Advantages

Smart and reliable volume adjustment

Single-handed volume setting

Interchangeable nozzle protection filter

Justip

adjustable tip ejector* fitting most tips

Swift-set user calibration system*

Shock, UV-light and autoclaving resistance

Pasteur pipette fitting 2 and 5mL models

* Socorex patented

Display AlwaysVisible

Located on the front of the instrument, the display window allows set volume to be visible

at all times during pipetting. A comforting reassurance for the user.

Smart and ReliableVolume Adjustment

State-of-the-art micrometric volume setting is performed smoothly and precisely by

turning push button. Finely machined click-stops and free rotating cap prevent any

unwanted volume alteration.

Patented Tip Ejection System

Ergonomically positioned, the large surface, soft padded ejector provides for low-pressure

activation. In addition, the shaft height adjustment system called Justip

allows optimal

distance setting between ejector and tip.

Pasteur Pipette Adapter

Adapter nozzles allow fitting of Pasteur pipette on Acura

®

manual

2 and 5mL models

in addition to polypropylene tips. The adapters have two O-rings guaranteeing proper

holding and tightness of standard 2mL Pasteur pipettes (Ø 7mm).

Nozzle Protection Filter

Protection against liquid overflow and contamination.

Inaccuracy (E%)

Imprecision (CV%)

Cat. No.

Volume

Division

Min. Vol.

Mid. Vol.

Max. Vol.

Min. Vol.

Mid. Vol.

Max. Vol.

Tip Style

W810315

0.2 - 2mL

0.002mL

< ± 1.5 % < ± 1.0 % < ± 0.5 %

< 0.5 % < 0.3 % < 0.2 %

2mL

W810316

0.5 - 5mL

0.01mL

< ± 1.5 % < ± 1.1 % < ± 0.6 %

< 0.6 % < 0.5 % < 0.3 %

5mL

W810318

1 - 10mL

0.01mL

< ± 1.5 % < ± 0.7 % < ± 0.5 %

< 0.5 % < 0.3 % < 0.2 %

10mL

Performance values obtained with double-distilled water at constant temperature (± 0.5°C) comprised between 20°C and 25°C in accordance with ISO 8655

.

Performance and Ordering Information - Acura

®

manual

835 Macropipettes

Fixed Volume Pipettes available at

www.wheaton.com

SOCOREX

®

Acura

®

manual

Pipettes

The Next Generation in Pipetting

835

0.2 - 2mL

0.5 - 5mL

1 - 10mL