Socorex Acura electro Pipettes - page 3

Cat. No.
Cat. No.
Inaccuracy (E%)
Imprecision (CV%)
Initial Package Pipette Alone
Volume
Division
Min. vol.
Mid. Vol.
Max. Vol.
Min. Vol.
Mid. Vol.
Max. Vol.
Tip Style
Micropipettes Acura
®
electro
926XS
0.1 - 2µL
0.01µL
<
±
2.5 %
1)
<± 1.2 % <± 0.9 % < 2.5 %
1)
< 1.5 % < 0.8 %
Ultra 10µL
0.5 - 10µL
0.05µL
<
±
1.2 %
2)
<± 0.8 % <± 0.6 % < 1.5 %
2)
< 0.7 % < 0.35 %
Ultra 10µL
0.5 - 10µL
0.05µL
<
±
1.2 %
2)
<± 0.8 % <± 0.6 % < 1.7 %
2)
< 0.8 % < 0.4 %
200µL
1 - 20µL
0.1µL
<
±
1.2 %
2)
<± 0.6 % <± 0.5 % < 1.2 %
2)
< 0.4 % < 0.3 %
200µL
2.5 - 50µL
0.25µL
<± 1.0 %
2)
<± 0.6 % <± 0.5 % < 0.7 %
2)
< 0.3 % < 0.25 %
200µL
5 - 100µL
0.5µL
<± 1.0 %
2)
<± 0.6 % <± 0.5 % < 0.7 %
2)
< 0.3 % < 0.2 %
200µL
10 - 200µL
1.0µL
<± 1.0 %
2)
<± 0.6 % <± 0.4 % < 0.6 %
2)
< 0.2 % < 0.15 %
200µL
50 - 1000µL
5.0µL
<± 0.8 %
2)
<± 0.5 % <± 0.4 % < 0.4 %
2)
< 0.15 % < 0.1 %
1000µL
Macropipettes Acura
®
electro
936
0.1 - 2mL
0.01mL
<± 1.5 %
2)
<± 1.0 % <± 0.5 % < 0.6 %
2)
< 0.3 % < 0.15 %
2mL
0.25 - 5mL
0.05mL
<± 1.2 %
2)
<± 0.8 % <± 0.5 % < 0.6 %
2)
< 0.3 % < 0.15 %
5mL
0.5 - 10mL
0.05mL
<± 1.0 %
2)
<± 0.7 % <± 0.5 % < 0.5 %
2)
< 0.2 % < 0.15 %
10mL
8-Channel Pipettes Acura
®
electro
956
0.5 - 10µL
0.05µL
<± 3.5 %
2)
<± 1.5 % <± 1.0 % < 3.0 %
2)
< 0.9 % < 0.7 %
Ultra 10µL
2.5 - 50µL
0.25µL
<± 1.0 %
2)
<± 0.9 % <± 0.8 % < 1.0 %
2)
< 0.6 % < 0.4 %
200µL
10 - 200µL
1.0µL
<± 0.9 %
2)
<± 0.7 % <± 0.6 % < 0.6 %
2)
< 0.4 % < 0.25 %
200µL
20 - 350µL
5.0µL
<± 1.0 %
2)
<± 0.8 % <± 0.6 % < 0.6 %
2)
< 0.4 % < 0.25 %
350µL
12-Channel Pipettes Acura
®
electro
956
0.5 - 10µL
0.05µL
<± 3.5 %
2)
<± 1.5 % <± 1.0 % < 3.0 %
2)
< 0.9 % < 0.7 %
Ultra 10µL
2.5 - 50µL
0.25µL
<± 1.0 %
2)
<± 0.9 % <± 0.8 % < 1.0 %
2)
< 0.6 % < 0.4 %
200µL
10 - 200µL
1.0µL
<± 0.9 %
2)
<± 0.7 % <± 0.6 % < 0.6 %
2)
< 0.4 % < 0.25 %
200µL
20 - 350µL
5.0µL
<± 1.0 %
2)
<± 0.8 % <± 0.6 % < 0.6 %
2)
< 0.4 % < 0.25 %
350µL
Performance values obtained in Forward mode with double-distilled water at constant temperature (± 0.5°C) comprised between 20 and 25°C in accordance with ISO 8655.
Performance measured at:
1)
0.5 µL,
2)
10% of nominal volume
Ordering Information
Initial package includes pipette, charging stand and accessories. Additional pipettes (alone) supplied with individual QC certificate, pipette tips samples and operating instructions.
Micropipettes
Reduced shaft length and conical end improve driveability
Easy access to microtubes and microplates
Optional longer volumetric modules
Simple conversion by user at any time
Unsurpassed metrological performance and stability
Macropipettes
Interchangeable nozzle protection filter
Adapters for Pasteur pipette and straw tips
Outstanding accuracy and precision
Multichannel
Large selection of 8 and 12-channel models
Lightweight and perfect hand fitting
Sequential tip ejection
Also fits any single channel volumetric module
A World of Pipetting Expertise with SOCOREX
®
Product Advantages and Instrument Performance
1,2 4,5,6